MG Work Vest Performance List

※Note

The protective performance of the MG work vest in this table is based on normal use in accordance with the handling instructions. Performance cannot be guaranteed if you intentionally approach something dangerous or if the MG work vest is not worn correctly.

※If you need something that is not listed in the table, please contact us.

Table

Things used in daily life ・Telecommunications Equipment ・Electrical appliances ・Other household appliances ・On the go
仕事場で使われるもの ・Telecommunications Equipment ・High Voltage Equipment
Medical Equipment ・Inspection Equipment ・Treatment Equipment

Symbols in the table

※Danger (dangers are indicated by symbols)
- No impact
! Correct use of the equipment is required.
Danger (pacemakers, ICDs, etc. may function improperly, but there is little risk of serious health problems)
Contraindications (There is a high possibility that pacemakers, ICDs, etc. may function improperly, which may result in serious health problems)
※Impact(Impact on pacemakers, ICDs, etc. are shown with symbols)
I Pacing is suppressed (the number of heartbeat signals from pacemakers, ICDs, etc. is reduced). It is necessary to leave the area as soon as possible.
T Unnecessary stimulation occurs. Demand function is lost, and conflict between the pacemaker, ICD, etc. and the patient's own pulse may occur. It is necessary to leave the area as soon as possible.
R The settings of the pacemaker, ICD, etc. will change and will not return to their original state. You will need to go to the hospital immediately and have the pacemaker, ICD, etc. reset.
D Pacemakers, ICDs, etc. may stop functioning or be destroyed.
※MG Work Vest Performance Symbols
- There is no need to wear an MG work vest.
Prevention of impact to pacemakers, ICDs, etc. has been confirmed.
By following our instructions, you can prevent impacts to pacemakers, ICDs, etc.
It has the ability to prevent impact on pacemakers, ICDs, etc., but the safety distance has not been confirmed.
The impact prevention performance on pacemakers, ICDs, etc. has not yet been confirmed.
× It is not possible to expect any impact prevention performance on pacemakers, ICDs, etc.
Telecommunications Equipment Electrical Products Other household appliances On the go
Workplace Communication Device WorkplaceHigh Voltage Equipment Medical Inspection Equipment Medical Treatment Equipment

Telecommunications Equipment

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
Mobile phone T/I Radio waves No impact if placed 15 cm or more away from pacemakers, ICDs, etc. ◎ Confirmed to prevent impact to pacemakers, ICDs, etc.
PHS / Cordless phone - - - No impact under normal use -
Bluetooth communication (PC devices) - - - -
ETC (Electronic Toll Collection) - - - -
ICOCA / Suica (JR fare systems) T/I Radio waves No impact if kept at least 12 cm from the IC card reader ◎ Confirmed to prevent impact to pacemakers, ICDs, etc.
Wireless LAN T/I Only some pacemakers and ICDs are affected. Check with the manufacturer.
Transceiver T/I Maintain a distance of approximately 30 cm from pacemakers, ICDs, etc.
On-board radio antenna (e.g. truck radio) T/I Impact may occur depending on the device and frequency No impact under normal use. Not covered for illegal radio waves.
Amateur radio station antenna T/I 〇〃
Remote control radios (e.g. model aircraft) T/I No impact as long as the antenna is not placed directly on the device

Electrical Products

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
Microwave oven T/I Radio waves / Current No impact unless directly held. Ensure proper grounding to avoid electrical leakage. ◎ Confirmed to prevent impact on pacemakers and ICDs. Caution: electrical leakage.
IH cooker (induction heater) T/I Magnetic field / Current Keep at least 50 cm away from pacemakers or ICDs. 〇 No impact unless the surface is covered or parallel to the stove.
Induction rice cooker T/I Keep at least 30 cm away from pacemakers or ICDs. 〇 No impact if not held during cooking.
Electric blanket / carpet T/I No impact unless in direct contact with pacemaker or ICD. Use of leakage-preventing type is recommended. - Use with MG Vest is recommended when operating equipment.
Electric massager T/I Generally no impact. Some models emit low-frequency waves or currents, so check specifications.
Electric hair dryer T/I No impact unless held near pacemaker or ICD. 〇 No impact if not pressed against the body.
Electric razor / electric toothbrush T/I No impact unless held near the chest. Rechargeable types are safer. 〇〃
Household power tools (e.g., drills, saws) T/I 〃 (Ensure proper grounding) 〇〃 (Ensure proper grounding)
Computers and peripherals T/I Current No impact. Caution: potential electrical leakage. - Ensure proper earthing and monitor for electrical leakage.
Television and remote control T/I -〃
Karaoke equipment T/I -〃
Home gaming devices T/I -〃
Refrigerator T/I -〃
Washing machine T/I -〃
Home electric sewing machine T/I -〃

Other household appliances

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
Body Fat Measuring Device T/I Current Do not use, as current passes directly through the body. ×
Low Frequency Therapy Device T/I/R ×
High Frequency Thermotherapy Device T/I/R Radio waves / Current
Electrical Therapy Device T/I/R Current ×
Magnet Static magnetic field No impact during normal use. Do not press directly against pacemakers or ICDs.
Magnetic Therapy Particles
Magnetic Bedding (mats, pillows, etc.)
Magnetic Necklace
Fully Automatic Mahjong Table T/I Magnetic field / Current Not recommended for individuals with pacemakers or ICDs. 〇 No impact unless the table surface is covered or operated in close contact.

On the go

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
Remote Controlled Golf Cart T/I Radio waves Possible impact due to radio waves from the remote control. 〇 No impact if not in direct contact with pacemakers or ICDs.
Metal Detectors (airports, facilities) T/I Magnetic field Generally safe, but pacemakers/ICDs may trigger detection; notify staff in advance. 〇 No impact if the gate is not touched directly.
Anti-theft Systems (libraries, electronics stores, etc.) T/I Do not linger near the gate; pass through quickly.
※ Only systems operating above 14kHz have been verified safe.
[See: How anti-theft devices affect pacemakers and ICDs]
Automotive Engines T/I/R Avoid standing near a running engine with the hood open. 〇 No impact except during starter motor operation.
Motorcycle/Tractor Engines (Gasoline) T/I/R Riding is generally discouraged; avoid running engines. 〇〃
Smart Key Systems T/I Radio waves / Magnetic field 〇 No impact during normal use.
Electric Vehicle Fast Chargers T/I Magnetic field Avoid use and do not approach without caution. ◎ Impact prevention for pacemakers/ICDs confirmed.
Linear Motor Train T/I/R Magnetic field / Static magnetic field No impact unless lying on the floor. 〇 Safe under normal usage conditions.

Workplace Communication Device

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
Radar Antenna T/I Radio waves Individuals with pacemakers or ICDs are prohibited from entering the facility. △ Protective effect confirmed, but safety verification is required.
Broadcasting Station Antenna T/I
Mobile Phone Base Station T/I

WorkplaceHigh Voltage Equipment

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
Arc Welding Equipment T/I Magnetic field Individuals with pacemakers or ICDs are not permitted to enter the facility. △ Protective capability confirmed, but safety verification is required.
Spot Welder T/I
High-Frequency Induction Melting Furnace T/I Radio waves
Large Motor T/I Magnetic field
Industrial Furnace T
Large Generator T/I
High-Voltage Equipment T

Medical Inspection Equipment

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
X-ray - - - No impact. However, be sure to inform the staff that you have a pacemaker or ICD. Remove the MG Vest before the procedure.
CT Scan - - - Impact may occur with certain pacemakers or ICDs. Please consult the device manufacturer.
PET - - - No impact. However, notify staff that you are using a pacemaker or ICD.
Bone Density Scanner - - -
Electrocardiogram (ECG) - - -
Ultrasound Echography - - -
Bioimpedance Measuring Instrument T/I Current Use only under guidance from a pacemaker or ICD specialist.
Other Current-Based Examinations T/I
MRI T/D Magnetic field / Radio waves Not suitable for individuals with pacemakers or ICDs.

Medical Treatment Equipment

Electromagnetic Impact Device danger Impact Cause of Impact Basic measures for those with pacemakers, ICDs, etc. MG Vest
※Performance/Notes
Laser Beam - - - No impact. However, notify medical personnel if you have a pacemaker or ICD. Remove the MG Vest prior to use.
Dental Electric Files/Drills - - -
Infrared Therapy - - -
Electromagnetic Shock Wave Lithotriptor T/I Current Use only under the guidance of a pacemaker/ICD specialist.
Electric Scalpel I/R/D Current
Electric Acupuncture Device (e.g., for lower back pain) T/I - Not suitable for use with pacemakers or ICDs.
External Defibrillator R/D Current
Radiation Therapy Equipment (γ-ray irradiation) D Gamma rays

Electromagnetic wave protection standards