Wearing and Maintenance

How to wear

The CES protector is worn over the examination gown (it can also be worn on bare skin).

To prevent electromagnetic waves from entering through gaps, the mask can be adjusted to fit your body size using Velcro or a belt.

  1. 1.First to put on
  2. 2.Overlap the rear hook-and-loop fastener
  3. 3.Adjust the size with hook-and-loop fastener
  4. 4.Belts on the sides to fit your body
  5. 5.Pull the cuffs to fit
  6. 6.Fasten the strings on the cuffs with stoppers
  7. 7.Fit it to your body and you're done!

Washing, care and storage instructions


Maintenance methods and quality assurance


Maintenance method

  • The outer material is waterproof and breathable nylon, and the lining is made of antibacterial, antibacterial, and deodorizing polyester. If the product becomes dirty or smells like sweat or oil after use, please use a disinfectant spray to keep it clean and maintain hygiene.
  • Low-temperature sterilization (gas sterilization and plasma sterilization) is possible.
  • If the item is severely soiled, dry clean it with a petroleum-based cleaner.
    Do not use a dehydrator or dryer.

quality assurance More information

  • The manufacturer's quality guarantee period is two years from the date of purchase (a copy of the warranty card and delivery slip is required).
  • If the MRI protector begins to wear down after two years, or if the electrical resistance of the electromagnetic shielding exceeds 10 Ω, it is time to replace it.

Please refer to the photo on the right:

  • The resistance value for this product's standards was determined by measuring the electrical resistance over a 10cm span of electromagnetic shielding fabric using a tester.
  • If the electrical resistance of the electromagnetic shielding cloth exceeds 10 Ω during use, the product is eligible for replacement.

※If the electrical resistance of the electromagnetic shielding fabric exceeds 10 Ω due to storage conditions, age, etc., it may not perform as intended.
Please note that if there are problems with storage conditions or harsh usage conditions, the product will not be eligible for replacement within the warranty period or without a warranty on usage conditions.

For safe use